Thermocouple check device SN / WJT-2A

Word document
  • File type: Word document
  • File size: 0.1M
  • Update: 2016-08-03 10:57
  • 下载次数: 327 Browse the number: 17066 Downloads the number: 327
Enter download
Detailed introduction
Thermocouple check device for industry

Model: SN / WJT-2A

WJT-2A thermocouple calibration device is specially used for calibration work, suitable for: nickel-chromium-nickel-silicon (K) nickel-chromium-silicon (N) nickel-chromium-copper nickel (constantan) (E) Calibration and indexing of thermocouples.
The device has a more advanced electrical measurement system and temperature control system, which has reliable verification data and stable performance.

The comparison method is adopted for verification: that is, the second-class standard platinum-rhodium 10-platinum thermocouple is directly compared with the calibrated thermocouple for verification and indexing.

The complete set of equipment includes second-class standard platinum-rhodium 10-platinum thermocouple, operation table, and thermocouple tubular verification furnace.

The operation console consists of a temperature control system and an electrical measurement system.

Temperature control system includes: XMTF intelligent digital regulator, temperature control thermocouple, solid state relay, etc.

The electrical measurement system includes: PZ150A DC digital voltmeter and SY821 transfer switch.

This device is suitable for provincial, city, and county-level legal metrological technical institutions, industrial and mining enterprises, and scientific research units.

Specifications

WJT-2A thermocouple calibration device is specially used for calibration work, suitable for: nickel-chromium-nickel-silicon (K) nickel-chromium-silicon (N) nickel-chromium-copper nickel (constantan) (E) Calibration and indexing of thermocouples

Product technical data

Working conditions:

Ambient temperature: 10 ~ 30 ℃

Relative humidity: ≤85 ﹪

Atmospheric pressure: 86 ~ 106Kpa

There is no corrosive gas and severe dust in the surrounding air except the geomagnetic field. There is no other external magnetic field around the device. The device is placed in a work place without vibration and shock.

Voltage power supply 220V ± 10 ﹪ 50HZ

Dimensions and weight:

Operating table (length × width × height) 1200 mm × 750 mm × 1250mm

Weight: about 160kg

Calibration furnace (length × width × height) 600 mm × 260 mm × 1130mm

Weight: 50kg

Technical Parameters:

1. 5 thermocouples tested

2. Temperature control range: 300 ~ 1200 ℃

3. Temperature control stability is not more than 0.2 ℃ / min

4. With automatic shutdown function of over temperature alarm

5. Electrical measurement system

PZ150A DC Digital Voltmeter

Measurement range: 1mV ~ 100.000mV

1uV

Basic error ± (0.005 ﹪ RD + 0.0015 ﹪ FS)

SY821 transfer switch

Parasitic potential

6. Temperature control system

XMTF-21 digital display regulator

Indexing number S type range 0 ~ 1600 ℃

Control mode intermittent PID

Temperature control thermocouple platinum rhodium 10-platinum thermocouple index number S type

Tube verification furnace

Working voltage 0 ~ 160V 50Hz

zui high use temperature 1200 ℃

The deviation of the center of the high temperature field from the furnace axis method is not greater than 10 mm.

Furnace temperature rise time from room temperature to 1100 ° C within 60 minutes

[ Information search ] [ ] [ Tell a friend ] [ Print this article ] [ Close the window ] [ Back to top ]

download link
  • Member Index:
  • Enterprise Certification:
  • Contact: Manager Gao (Ms.)
  • Phone: 010-59143021
  • Mobile: 15313875080
  • Region: Beijing
Recommended information
Top downloads in this category
Top downloads
七乐彩 欢乐拼三张 欢乐拼三张 七乐彩